• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH 5.0MM CANNULATED DRILL BIT LARGE QC/300MM; BIT,DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH 5.0MM CANNULATED DRILL BIT LARGE QC/300MM; BIT,DRILL Back to Search Results
Catalog Number 310.630
Device Problem Failure to Cut (2587)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that there was a wear and deterioration of the tips of two (2) screwdrivers and two (2) cannulated drill bits due to frequent use.The event occurred during internal service activities.There was no patient involvement reported.This report is for one (1) 5.0mm cannulated drill bit large qc/300mm this is report 3 of 4 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part number: 310.630, synthes lot number: pe02302, supplier lot number: n/a, release to warehouse date: 04sep2015, expiration date: n/a, supplier: (b)(4).No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Visual investigation: the received device is in a well-used and worn condition.Especially the drill bit section at the tip are worn (rounded and dull cutting edges) from use.Investigation conclusion: after a visual inspection per guidance, it is determined that the reusable instrument is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.End of life definition per important information leaflet, limits on reprocessing: end of life of a device is normally determined by wear and damage due to use.Evidence of damage and wear on a device may include but is not limited to corrosion (i.E.Rust, pitting), discoloration, excessive scratches, flaking, wear and cracks.Improperly functioning devices, devices with unrecognizable markings, missing or removed (buffed off) part numbers, damaged and excessively worn devices should not be used.No manufacturing related issue was identified and/or confirmed.Based on the investigation findings, it has been determined that no corrective and/or preventative action is appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0MM CANNULATED DRILL BIT LARGE QC/300MM
Type of Device
BIT,DRILL
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9734586
MDR Text Key192048730
Report Number8030965-2020-01286
Device Sequence Number1
Product Code HTW
UDI-Device Identifier07611819065192
UDI-Public(01)07611819065192
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number310.630
Device Lot NumberPE02302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2020
Date Manufacturer Received04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-