The reported failure was already investigated under complaint # (b)(4) with the adhesive membrane of the pven pressure sensor has come loose.The adhesive membrane of the pint pressure sensor is yellowish and looks hardened.Hls module connected to the cardiohelp.All values remain constant.The sensors became wet and as a result the sensors can lead to erroneous measurements.Thus the reported failure could be confirmed.The most probable root cause is that the sensors got wet and therefore displayed wrong pressures.When the event occurred, the device was being used for treatment of the patient.The product was directly involved in the incident.The occurrence rate regarding the above complaint is below the acceptance rate.Thus, no remedial action required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|