OBERDORF SYNTHES PRODUKTIONS GMBH MATRIXMANDIBLE SHORT CUT PLATE CUTTER; INSTRUMENT,CUTTING,ORTHOPAEDIC Back to Search Results
Catalog Number 03.503.057
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from (b)(6) reports an event as follows: it was reported that on an unknown date that the short cut plate cutters (03.503.057) were broken during a reconstruction surgery for a mandibular.The surgery was completed by using a trauma plate with a 30 minute surgical delay.Concomitant device: unknown plate (part # unknown, lot # unknown, quantity 1).This report is for one matrixmandible short cut plate cutter.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part: 03.503.057, lot: 8585345, manufacturing site: hägendorf, release to warehouse date: 25.November 2013.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: the visual inspection of the shortcut has shown wear marks, scratches and strongly damages on the cutting area of the instrument.Investigation conclusion: after a visual inspection per guidance, it is determined that the reusable instrument is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.End of life definition per important information leaflet, limits on reprocessing: end of life of a device is normally determined by wear and damage due to use.Evidence of damage and wear on a device may include but is not limited to corrosion (i.E.Rust, pitting), discoloration, excessive scratches, flaking, wear and cracks.Improperly functioning devices, devices with unrecognizable markings, missing or removed (buffed off) part numbers, damaged and excessively worn devices should not be used.No manufacturing related issue was identified and/or confirmed.Based on the investigation findings, it has been determined that no corrective and/or preventative action is appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIXMANDIBLE SHORT CUT PLATE CUTTER
Type of Device
INSTRUMENT,CUTTING,ORTHOPAEDIC
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9738479
MDR Text Key191587502
Report Number8030965-2020-01291
Device Sequence Number1
Product Code HTZ
UDI-Device Identifier07611819894426
UDI-Public(01)07611819894426
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.503.057
Device Lot Number8585345
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2020
Date Manufacturer Received03/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SHORTCUT F/MATMAND-PL T1.5-2.8 W/RASP IN; SHORTCUT F/MATMAND-PL T1.5-2.8 W/RASP IN; TIPREMED RECON PL/SM/RT/2.5MM THK-ST
-
-