• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PRSTHSIS,HIP,SM-CONSTRAINED, UNCMNTD,MAL/PLYER,NON-PROS, CLICUM-PHSPHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY INSERTER POSTER HARD; PRSTHSIS,HIP,SM-CONSTRAINED, UNCMNTD,MAL/PLYER,NON-PROS, CLICUM-PHSPHT Back to Search Results
Model Number 71365705
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2020
Event Type  malfunction  
Event Description
It was reported during surgery the inserted broke while inserting the implant.A backup device was available to complete the procedure.
 
Manufacturer Narrative
The device, used in treatment was not returned for evaluation, reporting event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY INSERTER POSTER HARD
Type of Device
PRSTHSIS,HIP,SM-CONSTRAINED, UNCMNTD,MAL/PLYER,NON-PROS, CLICUM-PHSPHT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9738972
MDR Text Key181384206
Report Number1020279-2020-00653
Device Sequence Number1
Product Code MEH
UDI-Device Identifier03596010555663
UDI-Public03596010555663
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71365705
Device Catalogue Number71365705
Date Manufacturer Received02/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-