The results from the in-house retain testing (sensor configuration 35.N) were found to be normal and well within product specifications with no observable irregularities in testing or during data analysis.The epoc potassium performance on card lot 10-19324-30 was deemed to be acceptable, with no product deficiencies identified.The cause of the discrepancy could not be determined from the information provided by the customer, but there is some suggestion that the sample presented to the epoc system may have had clotting or mixing related issues.The failure rate of lot 10-19324-30 is not showing an increased trend in the field, therefore there is no further evidence that the system or reagent cards are not performing as intended.
|