• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ULTRA AGGRESSIVE PLUS 4.0MM 5PK ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ULTRA AGGRESSIVE PLUS 4.0MM 5PK ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE Back to Search Results
Model Number 283419
Device Problem Break (1069)
Patient Problem Not Applicable (3189)
Event Date 02/10/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint#: (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Udi: (b)(4).
 
Event Description
It was reported by the affiliate via complaint submission tool that during an unknown procedure 8 of the aggressive 4. 0mm 5pk released metallic debris in the joint. There was no patient consequence, however there was a 5 minutes surgical delay. The procedure was completed washing the joint. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameULTRA AGGRESSIVE PLUS 4.0MM 5PK
Type of DeviceARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key9744753
MDR Text Key192160557
Report Number1221934-2020-00658
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K041824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/10/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number283419
Device Catalogue Number283429
Device Lot NumberM1906037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage

-
-