Model Number N/A
Device Problems Material Frayed (1262); Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/31/2020
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformance's related to the reported event were noted. The facility's biomedical engineer (biomed) ordered and replaced the power cord (0012-00-1688-21). The biomed indicated that the power cables were extended on all their cardiosave's to prevent the issue from reoccurring. The iabp was cleared for clinical use. Not returned to manufacturer.
 
Event Description
It was reported that during use on a patient, the power cord would not extend from the cardiosave intra-aortic balloon pump (iabp). The facility's biomedical engineer found that the cord had wrapped internally around one of the guide post and the insulation had been damaged in the process. There was no harm or injury to patient and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key9751603
MDR Text Key192193319
Report Number2249723-2020-00319
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/25/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/04/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-