WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS GAUGE, DEPTH Back to Search Results
Model Number 319.006
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Additional narrative: device returned. A review of the device history record. Part # 319. 006 synthes lot # h299834 supplier lot # na release to warehouse date: 21 oct 2017. Manufactured by synthes monument. No ncr's were generated during production. Device history batch null, device history review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition. Investigation summary updated event description. It was reported that on an unknown date during routine incoming inspection, the depth gauge was bent. There was no patient involvement. Investigation flow: damage visual inspection: visual inspection performed at customer quality (cq) showed the needle is not bent; instead it broke off at juncture between the needle and slider. No other issues were identified. A device failure was identified. Dimensional inspection: based on the review of the above drawings, no design issues contributing to relevant complaint condition were identified. Conclusion: no definitive root cause was able to be determined as the circumstances surrounding the complaint are unknown. During the investigation no unidentified product design/manufacturing issues or discrepancies were observed that may have contributed to the complaint condition. Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date during routine incoming inspection, the depth gauge was bent. There was no patient involvement. This complaint involves one (1) device. This report is 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of DeviceGAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key9751865
MDR Text Key191179153
Report Number2939274-2020-01049
Device Sequence Number1
Product Code HTJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 02/04/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number319.006
Device Catalogue Number319.006
Device Lot NumberH299834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2020
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Unkown

-
-