MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 25feb2020.
 
Event Description
It was reported that the ventilator logged a message of primary audio failed.There was no patient involvement.The service engineer (se) inspected the device.The se found a malfunction in the speakers.The se replaced the speakers to address the reported issue and the unit passed all tests.
 
Manufacturer Narrative
G4: 03jun2020.B4: 04jun2020.A main speakers were returned for analysis.A visual inspection of the returned components was performed, no notable conditions were found.An investigation was performed and the product analysis technician reported that the root cause was isolated to the primary speaker labeled #1 voice coil which was found to be open with 983.27 kilo-ohms the primary speaker labeled #2 was tested with no failures were found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9753349
MDR Text Key184287638
Report Number2031642-2020-00619
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-