Model Number 1MTEC30 |
Device Problem
Failure to Eject (4010)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.Lot number: unknown, information not provided.Unique device identifier (udi #): complete udi# is unknown, as lot number was not provided.Expiration date: unknown, as lot number was not provided.If implanted, give date: not applicable as cartrdige is not an implantable device. if explanted, give date: not applicable as cartrdige is not an implantable device. device manufacture date: unknown as lot number was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Event Description
|
It was reported that during insertion of the intraocular lens (iol) there was a loading issue, and lens was stuck in cartridge, would not advance.There was patient contact with the eye but no medical or surgical intervention were required.Patient is doing fine post-op.No further information provided.
|
|
Manufacturer Narrative
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
|
Manufacturer Narrative
|
Device evaluation: the reported cartridge was not returned for investigation.Therefore, the sample evaluation could not be performed, and the reported issue could not be verified.Product deficiency could not be identified.Manufacturing record evaluation: manufacturing record review cannot be performed since the lot number is unknown.A search of complaints related to lot number cannot be performed since the lot number is unknown.Conclusion: since no sample was returned and the lot number is unknown an investigation could not be performed, and no malfunction is confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|