MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97714
Device Problems Display or Visual Feedback Problem (1184); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978); Operating System Becomes Nonfunctional (2996)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 02/11/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 37761, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient.It was reported that there was a loose connector pin on the desktop charger.On the date of this report, the patient plugged the recharger into the wall to charge and later when they went to charge their implantable neurostimulator (ins), they noticed that the recharger screen was blank and the device was unresponsive.The patient stated that the green light on the desktop charger was on and that they had tried multiple outlets but the recharger was not taking a charge.Since the patient could not charge their ins, they now had a return of symptoms.Troubleshooting steps were performed during the call and the patient noticed that the connector pin was wiggly.The repair department was contacted and a replacement desktop charger was requested.No further complications were reported or anticipated.Indication for use is spinal pain.
 
Manufacturer Narrative
Analysis of the desktop charger (serial# (b)(4)) found bent connector pins.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9759420
MDR Text Key181337574
Report Number3004209178-2020-04154
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2018
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2020
Date Manufacturer Received03/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age43 YR
-
-