Partial information about the patient has been provided.
Livanova (b)(6) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.
The z number is z-2076/2081-2015.
Livanova (b)(6) manufactures the heater-cooler system 3t.
The incident occurred in (b)(6).
Through further follow-up communication with the chief perfusionist livanova learned that the water in the heater-cooler systems 3t in use is changed every day and they are stored dry.
This is not in alignment with current instruction for use however reportedly the devices are very clean and there is no sign of biofilm.
The devices are located inside the operating theatre during use.
We are currently waiting to know the result of microbial sampling performed at customer site.
Reportedly the tests are still in progress.
If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|