• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Microbial Contamination of Device (2303)
Patient Problem Bacterial Infection (1735)
Event Date 07/16/2019
Event Type  Injury  
Manufacturer Narrative
Partial information about the patient has been provided. Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices. The z number is z-2076/2081-2015. Livanova deutschland manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Through further follow-up communication with the chief perfusionist livanova learned that the water in the heater-cooler systems 3t in use is changed every day and they are stored dry. This is not in alignment with current instruction for use however reportedly the devices are very clean and there is no sign of biofilm. The devices are located inside the operating theatre during use. We are currently waiting to know the result of microbial sampling performed at customer site. Reportedly the tests are still in progress. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a female patient (date of birth (b)(6)undergone cardiac surgery on (b)(6) 2019 and an heater-cooler system 3t was used. On (b)(6) 2019 the patient was confirmed to be infected with mycobacterium avium intracellulare complex and mycobacterium chimaera was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9764310
MDR Text Key192681217
Report Number9611109-2020-00146
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/27/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number16-02-85
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/01/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured05/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015

-
-