Model Number C408646
Device Problem Gas/Air Leak (2946)
Patient Problems Air Embolism (1697); Arrhythmia (1721)
Event Date 02/11/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During the procedure, the patient experienced st elevation and a heart block.Right after blood was purged from the catheter, there was transient, diminished flow that settled with nitrates inside the vessel.It was alleged that there was a microbubble from the catheter despite the catheter being primed with contrast outside the patient.The av block was diagnosed using an ecg monitor.Medication was administered to stabilize the patient.(patient id: (b)(6), clinical study: (b)(6)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY OPTIS KIT BOX IMAGING CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
Manufacturer (Section G)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key9768987
MDR Text Key188861714
Report Number3009600098-2020-00007
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K141453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Model NumberC408646
Device Catalogue NumberC408646
Device Lot Number7278523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-