A steris service technician and the user facility's biomed department inspected the lighting system and did not observe any particulates falling during their inspection.During the inspection, the technician observed paint chipping on the suspension components which is indicative of impact damage as well as residual white material, dried blood, and dust on the lighting system.The technician replaced the suspension arm due to the observed paint chipping, tested the light system, confirmed it to be operating according to specification, and returned it to service.The harmonyair m series surgical lighting system operator manual states (1-3), "caution - possible equipment damage: do not bump lightheads into walls or other equipment.Always use handles or gripping surface when positioning lighthead during surgical procedures, or when cleaning or servicing the lighting system." the operator manual further states (6-4), "the following areas of the lighting system must be cleaned and disinfected before each use of the system: lighthead and monitor suspension arms, lighthead, lighthandle" the account manager offered in-service training on the proper use, operation, and cleaning of the harmonyair m series surgical lighting system; however, the user facility declined.No additional issues have been reported.
|