• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9445-02
Device Problem No Device Output (1435)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/16/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that transmitter failed error occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that transmitter failed error occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage test was performed and failed due to 0v.No data was provided for evaluation.Confirmation of the allegation was undetermined.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key9772535
MDR Text Key181531448
Report Number3004753838-2020-020182
Device Sequence Number1
Product Code QBJ
UDI-Device Identifier00386270000385
UDI-Public00386270000385
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9445-02
Device Catalogue NumberSTT-OR-001
Device Lot Number5265091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age10 YR
Patient Weight52
-
-