• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD LEGACY 1 PUMP; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD LEGACY 1 PUMP; PUMP, INFUSION Back to Search Results
Model Number 6400
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problems No Patient Involvement (2645); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
One cadd legacy 1 pump was returned for evaluation.The event history log was reviewed and there was no evidence of the reported issue.Three separate delivery accuracy tests were conducted and the pump was visually inspected.The reported failed accuracy was unable to be confirmed.Delivery accuracy testing found the pumps average delivery error to be within the published specification of +/-6%.There was no fault found with the returned pump.
 
Event Description
Information was received indicating that a smiths medical cadd legacy 1 pump failed accuracy by +6.75%.There was no reported adverse event.
 
Manufacturer Narrative
Additional information: h6, h10.Additional information was received that this issue occurred while testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD LEGACY 1 PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis, mn
MDR Report Key9772565
MDR Text Key181537957
Report Number3012307300-2020-01776
Device Sequence Number1
Product Code FRN
UDI-Device Identifier10610586019548
UDI-Public10610586019548
Combination Product (y/n)N
PMA/PMN Number
K982838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor
Type of Report Initial,Followup
Report Date 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number6400
Device Catalogue Number21-6400-51
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2020
Date Manufacturer Received03/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-