• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO NC SPRINTER RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO NC SPRINTER RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Device Problems Deflation Problem (1149); Fracture (1260)
Patient Problem Injury (2348)
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
An nc sprinter rx ptca balloon catheter was attempted to be used.Negative prep was performed with a syringe and the device was directly connected to a non medtronic indeflator.It was reported that the luer of the balloon cracked during the procedure.It was reported that the balloon was inflated to 4-6 atms and then deflation issues were encountered.An attempt was made to pull an additional vacuum using a 30cc syringe but there was still difficulty getting the balloon down.The system was eventually retracted.There was no injury to the patient.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC SPRINTER RX
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key9785045
MDR Text Key181920462
Report Number9612164-2020-00995
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P790017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-