OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.Aeromonas hydrophila and aeromonas caviae (the total cfu of the aeromonas hydrophila and aeromonas caviae is 1 cfu/endoscope).The device had been reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus france (ofr).Ofr sent the device to a third party laboratory for microbiological testing.As a result of the testing, the negligible numbers of microbes were detected on the sample collected from the all channels of the subject device, but the testing result cleared the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key9785155
MDR Text Key194489909
Report Number8010047-2020-01602
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 05/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-