Model Number URF-P7 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.Aeromonas hydrophila and aeromonas caviae (the total cfu of the aeromonas hydrophila and aeromonas caviae is 1 cfu/endoscope).The device had been reprocessed using peracetic acid.There was no report of infection associated with this report.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus france (ofr).Ofr sent the device to a third party laboratory for microbiological testing.As a result of the testing, the negligible numbers of microbes were detected on the sample collected from the all channels of the subject device, but the testing result cleared the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
|
|
Search Alerts/Recalls
|