MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX EMBOLIZATION DEVICE INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-375-14
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/28/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the pipeline pushwire broke during retrieval. The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm in the internal carotid artery with a max diameter of 4mm and a 2. 5mm neck diameter. It was noted the patient's vessel tortuosity was moderate. It was reported that during deployment of the pipeline, the healthcare provider (hcp) attempted to recapture the delivery system. While trying to pull the wire into the catheter, it separated at the proximal bumper. The hcp recognized this and removed the delivery wire and the portion that broke free. There was no issue reported with the catheter used, and the post procedure angiographic results were fine. It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event. Ancillary devices include a phenom catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePIPELINE FLEX EMBOLIZATION DEVICE
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
katcha taylor
9775 toledo way
irvine, CA 92618
9496801345
MDR Report Key9785499
MDR Text Key196738599
Report Number2029214-2020-00192
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100018.S011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/27/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/14/2022
Device Model NumberPED-375-14
Device Catalogue NumberPED-375-14
Device Lot NumberA815066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-