• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT40IDX56OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT40IDX56OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS Back to Search Results
Model Number 1218-87-456
Device Problems Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Edema (1820); Foreign Body Reaction (1868); Nerve Damage (1979); Tissue Damage (2104); Discomfort (2330); No Code Available (3191)
Event Date 02/19/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient was unstable.Doi: (b)(6) 2009; dor: (b)(6) 2020; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6 patient code: no code available (3191) used to capture the patient code blood heavy metal increased.
 
Event Description
Pinnacle claim submission form and medical records received.After review of medical records, the patient was revised to address metal on metal wear.There was gray-tinged fluid within the joint, corrosion on the femoral head-trunnion junction and only 75% of the abductor muscle's insertion was intact with some evidence of denervation and trunnionosis.Lab results note fluid collection, capsular dehiscence, an elevated cobalt level, discomfort and possible adverse tissue reaction to metal-on-metal bearing.Doi: (b)(6) 2009 - dor: (b)(6) 2020 (right hip).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint # (b)(4).Investigation summary - no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot - null.Device history batch - null.Device history review - null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE MTL INS NEUT40IDX56OD
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9799909
MDR Text Key187557090
Report Number1818910-2020-07083
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K062426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2014
Device Model Number1218-87-456
Device Catalogue Number121887456
Device Lot Number2846464
Was Device Available for Evaluation? No
Date Manufacturer Received04/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
12/14 ARTICUL 40MM M SPEC+5.; CUP 121722056/DN7ET1000.; SCREW 121725500/B4KCW4000.; SCREW 121730500/C78FN4000.; STEM 157011120/DJ4A11.
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient Weight67
-
-