• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MODULAR CALCAR PLANER SML; REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MODULAR CALCAR PLANER SML; REAMERS Back to Search Results
Model Number 2570-04-501
Device Problems Dull, Blunt (2407); Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/19/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported  that the srom hudson adapter broke where it attaches to instrument.40mm and 48mm calcar planars are dull and would not ream away bone adequately.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary : the device associated with this report was not returned, thus the reported event could not be confirmed.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.H10 additional narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR CALCAR PLANER SML
Type of Device
REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9804964
MDR Text Key190814692
Report Number1818910-2020-07153
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295482024
UDI-Public10603295482024
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2570-04-501
Device Catalogue Number257004501
Device Lot NumberPG272842
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2020
Date Manufacturer Received07/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age70 YR
-
-