Related manufacturer reference number: 1627487-2020-02553; related manufacturer reference number: 1627487-2020-02554; related manufacturer reference number: 3006705815-2020-01089.It was reported that the patient¿s stimulation kept stopping on its own.Additionally, patient reported sometimes experiencing electrical sensations at the lead/extension connection site.Diagnostics revealed high impedances.X-ray revealed that one of the lead tails was not completely inserted into the extension.Reprogramming was unable to resolve the issue.As such, surgical intervention took place on (b)(6) 2020.During intra-op, impedance reading was normal on the lead and extension.However, the impedance reading was high when measured on the entire system.The extensions were replaced with new extensions, but the issue persisted.In turn, the ipg was also explanted and replaced with a ipg to address the issue.No intervention took place on the lead.
|