MEDTRONIC NAVIGATION, INC SURGN CART 9735665 STEALTH S8 PREMIUM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9735665
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/27/2020
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9735670, serial/lot #: (b)(4), udi#: (b)(4); product id: 9735796, serial/lot #: unknown.Onsite functional and visual examination was performed by a manufacturer representative.Hardware parts were replaced and the issue was resolved.The system passed a system checkout and was returned to an operational condition.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system used during a sacroiliac and thoracolumbar procedure.It was reported that the site's system during a clinical case was noted to power cycle on the camera cart pcoip screen.There was no allotted delay in the case and patient was present during the complaint.
 
Manufacturer Narrative
D4: lot number for pcoip provided h3: the pcoip was returned to the manufacturer for analysis.The issue was confirmed.Functional testing determined that pcoip client was tested and logs indicated 1 software reboot.Configured to auto negotiation.The failure mechanism determined defective pcba codes associated with the pcoip: fdr 120, fdc 4307 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGN CART 9735665 STEALTH S8 PREMIUM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9810742
MDR Text Key183327156
Report Number1723170-2020-00815
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Device Lot Number71AB42C6J2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2020
Date Manufacturer Received03/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age71 YR
Patient Weight104
-
-