Model Number 9300TFX26J
Device Problems Failure to Conduct (1114); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Arrhythmia (1721); No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2016
Event Type  Injury  
Manufacturer Narrative
Udi number: (b)(4).Per the instructions for use (ifu), conduction system defects (heart block) which may require a permanent pacemaker are potential adverse events associated with balloon aortic valvuloplasty, deployment of the prosthetic valve, and the overall tavr procedure.According to the valve academic research consortium (varc) guidelines, the close anatomical relationship between the aortic valve complex and the branching atrioventricular bundle may provide an explanation for these complications of the tavr procedure.According to literature review, and as documented in a technical summary written by edwards lifesciences, atrioventricular conduction disturbances after tavr are associated with many patient related and procedural related factors, including pre-operative co-morbid status, the degree and bulkiness of aortic valve and annular calcification, inter-ventricular septal thickness, pre-existing electrocardiogram abnormalities, the depth of prosthesis implantation, and the profile of the implanted prosthesis.Unlike conventional avr, where there may be localized trauma due to decalcification of the annulus and/or suture placement in the proximity of the av node or the bundles, tavr may cause conduction abnormalities through mechanical impingement of the conduction system by the prosthesis.The mechanisms of the development of heart block after tavr are well documented and described in the literature.It is also documented that pre-existing heart block is common in patients undergoing tavr or surgical avr and another 4-6 % will develop postoperative heart block, potentially requiring a permanent pacemaker.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.Patient factors not provided and/or the mechanisms described above are likely contributing factors for the event.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Event Description
As reported through the (b)(6) registry by our affiliates in (b)(6), post implant of a 26mm sapien xt valve in the aortic position, the patient developed a conduction disturbance.Permanent pacemaker implantation was performed on postoperative day (pod) 3.The patient¿s outcome was determined to be ¿recovered¿.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE (THV)
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key9814181
MDR Text Key190816405
Report Number2015691-2020-10878
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
PMA/PMN Number
P130009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/23/2018
Device Model Number9300TFX26J
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-