The subject device has not been returned to omsc for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be determined, but it is surmised that the hole in the rubber of the bending section occurred as the cable support detached from the bending section of the subject device.In addition, it is surmised that the bending section was broken and the metal was stuck out from the bending section because the user facility pushed the device while it is angulated.
|