• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problems Use of Device Problem (1670); Malposition of Device (2616)
Patient Problems Foreign Body Reaction (1868); Pain (1994); No Code Available (3191)
Event Date 01/07/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4). Initial reporter occupation: lawyer. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
After review of medical records, it was indicated that the patient had right hip pain with metallosis. The patient was then revised for failed right tha. Operative notes reported metallosis containing tissue was excised. Operative finding reported thickening of the greater trochanteric bursa with some black stained tissue. Trunnions. Stem with 20 degree anteversion. Cup with 45 degree abduction and 25 degrees anteversion. Doi: (b)(6) 2004; dor: (b)(6) 2019; right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN HIP ACETABULAR CUP
Type of DeviceHIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9819179
MDR Text Key182986744
Report Number1818910-2020-07494
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/11/2020 Patient Sequence Number: 1
-
-