Model Number 25-40-45 |
Device Problem
Gas Output Problem (1266)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.A livanova field service representative was dispatched to the facility to investigate the device and could not duplicate the reported error code and another error associated to a discrepancy between the actual and set values popped up at the start up.He noticed that the device was supplying variable air output even if it was set to zero.Investigation is still ongoing.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 gas blender system was giving an error code associated to an ad tranformer fault during setup.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: no further investigation is possible for this specific case since the device was not made available for investigation.The perfusionist informed livanova that they will discontinue the use if this affected blender and continue to use another device.The device will not be repaired.
|
|
Event Description
|
See initial report.
|
|