Model Number IPN000320 |
Device Problems
Battery Problem (2885); Power Problem (3010)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the intra-aortic balloon pump (iabp) would keep switching back and forth from ac power to battery power.The power cord was checked to be secure and the iabp was plugged into multiple different outlets with no change.The patient remained supported by the iabp with no disruption.As a result, the iabp was swapped out.There was no report of patient complications, serious injury or death.
|
|
Event Description
|
It was reported that the intra-aortic balloon pump (iabp) would keep switching back and forth from ac power to battery power.The power cord was checked to be secure and the iabp was plugged into multiple different outlets with no change.The patient remained supported by the iabp with no disruption.As a result, the iabp was swapped out.There was no report of patient complications, serious injury or death.
|
|
Manufacturer Narrative
|
Qn#(b)(4).No iap part was returned to teleflex chelmsford for investigation.The reported complaint of the pump keeps switching back and forth from ac power to battery power is not able to be confirmed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the iap serial and lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
|