Model Number 1217-22-054
Device Problems Difficult to Remove (1528); Use of Device Problem (1670); Malposition of Device (2616)
Patient Problems Foreign Body Reaction (1868); Pain (1994); No Code Available (3191)
Event Date 06/01/2018
Event Type  Injury  
Manufacturer Narrative

(b)(4). Initial reporter occupation: lawyer. (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.

 
Event Description

Pinnacle claim submission form and medical records received. After review of medical records the patient was revised to address pain and periprosthetic femoral fracture. Revision notes reported a loose femoral stem, black pigmentation at the fracture site suggestive of metallosis and pathological fracture, vertical and retroverted cup, metallosis in the cup, and a jammed liner; the whole acetabular components were removed. Doi: (b)(6) 2009; dor: (b)(6) 2018; right hip.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePINNACLE SECTOR II CUP 54MM
Type of DevicePINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9829101
MDR Text Key183272378
Report Number1818910-2020-07753
Device Sequence Number1
Product Code KWA
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK073504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER,OTHER
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/25/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received03/13/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date05/31/2018
Device MODEL Number1217-22-054
Device Catalogue Number121722054
Device LOT NumberCT2GE1000
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/22/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured05/16/2008
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 03/13/2020 Patient Sequence Number: 1
-
-