Model Number 2420-0007 |
Device Problem
Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
|
Event Description
|
The reported feedback suggests that the spike is broken.From the reported information there are no indications of serious injury to the patient or user as a result of this incident.
|
|
Event Description
|
The reported feedback suggests that the spike is broken.
|
|
Manufacturer Narrative
|
The customer's report that the spike was broken was confirmed.A visual inspection confirmed the customer¿s report as the tip of the spike of the 2420-0007 had snapped and was lodged within the spike port adaptor of the 82113eh.The root cause of the customer' report could not be determined.
|
|