The pipeline flex pushwire was returned for evaluation.
There was no pipeline flex braid returned with the pushwire.
The pipeline flex pushwire was found outside of the phenom 27 catheter.
The pipeline flex pushwire appeared to be detached at the hypotube proximal to the wire weld.
No defects were found with the tip coil, distal marker, re-sheathing marker or with the proximal bumper.
The ends of the detached pushwire were sent out for sem/eds analysis.
Based on the analysis findings and sem/eds analyses, the pipeline flex was confirmed to have "pushwire detached at hypotube proximal to the wire weld".
A review of the manufacturing process and lot history record did not uncover any deficiencies.
In addition, the elemental analysis conducted through scanning electron micrographic (sem) / energy dispersive spectroscopy (eds) showed presence of soldering material (tin); thereby indicating that the soldering was conducted.
If information is provided in the future, a supplemental report will be issued.
|