MEDTRONIC NAVIGATION, INC STEALTH S8 EM CRANIAL; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9735668 |
Device Problem
Imprecision (1307)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.Other relevant device(s) are: product id: 9735736, version: 1.2.0.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used for a functional endoscopic sinus surgery (fess) procedure.It was reported that there was a 3-8 mm lower and forward inaccuracy noticed immediately prior to navigation.It was noted that the site used tracer registration and the inaccuracy was being seen with multiple instruments.Two instrument sets were tested and an instrument problem was excluded.The location of the reference frame relative to the emitter tracking window was right and closer to middle point and the location of the reference frame relative to the patient was upper and middle of forehead.It was noted that the crosshair on the navigation was lower and forward than the instrument tip.In this account restarting registration resolved the problem, but the issue was still not resolved at the time of report.There was a one hour or longer surgical delay and no impact on patient outcome.
|
|
Manufacturer Narrative
|
Software analysis was completed by medtronic.Archives were reviewed.Registration was done with 1.3mm accuracy.The patient had a high bmi and a lot of loose skin.Some trace points were underneath the skin due to high skin pressure during collection of trace points.There is not enough evidence to determine a root-cause of the issue.Fda codes 10, 213, and 4315 are relevant to the software analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
D2 correction) product code updated from haw to pgw.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|