There was no patient involvement.
The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).
Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.
The z number is z-2076/2081-2015.
Livanova (b)(4) manufactures the heater-cooler system 3t.
The incident occurred in (b)(6).
A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
The laboratory report confirming the reported contamination has been provided to livanova.
Through follow-up communication livanova learned that the device was cleaned regularly per the instruction for use and that it was placed inside the operating theatre during use with the fan away from the patient and at an estimated distance of two (2) meters from the surgery field.
If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|