• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION Back to Search Results
Model Number PE075F5
Device Problem Pacing Problem (1439)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/06/2020
Event Type  Malfunction  
Manufacturer Narrative

No product was returned for evaluation; it was discarded at the hospital. Without the return of the product, it is not possible to determine if damages or defects existed on the product, nor could a root cause or potential contributing factors be identified. A device history record review was completed and documented that device met all specifications upon distribution. No actions will be taken at this time. Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients. There are multiple failure modes that may require the exchange of a pacing catheter. Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter. Stretching, kinking, or forceful wiping of the catheter may result in damage. After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration. Care should be taken not to kink the catheter body when securing it. In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event. Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.

 
Event Description

It was reported that the swan ganz catheter was unable to pace from the beginning of use. The catheter was replaced and the problem was solved. It is unknown if the patient had cardiac conduction defect. Patient demographic information requested but unavailable. There were no patient complications reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of DevicePACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn thomas
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key9861031
MDR Text Key188657471
Report Number2015691-2020-11056
Device Sequence Number1
Product Code LDF
Combination Product (Y/N)N
Reporter Country CodeJA
PMA/PMN NumberK813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,DISTRI
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial
Report Date 03/06/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received03/20/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberPE075F5
Device LOT Number62661430
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/06/2020
Was Device Evaluated By Manufacturer? No
Date Device Manufactured10/24/2019
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-