Catalog Number 316-3174-K01
Device Problem Inability to Irrigate (1337)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that irrigation did not flow during a procedure.The product was replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
Additional information provided in d.10., h.3., h.6., and h.10.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The wet returned sample was visually inspected and no obvious defects were observed that would have contributed to the reported event.A calibrated infiniti console was used to test the sample and the cassette primed and tuned with the ultrasonic handpiece successfully and could achieve maximum vacuum.No system message was generated, no fluid or air leaks, and no cracks were observed on the connectors that would have contributed to the reported event.The irrigation and aspiration flow rates were measured and found to be within specifications.Fluid flowed from the balance salt solution bottle to the irrigation and aspiration manifolds and continuously to the cassette housing.No occlusion or obstruction was observed during inspection and functional testing.The root cause of the customer's complaint could not be established as the returned cassette was evaluated and met specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key9866075
MDR Text Key186532611
Report Number1644019-2020-00169
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 06/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Catalogue Number316-3174-K01
Device Lot Number2290512H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2020
Date Manufacturer Received05/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED CUSTOM PAK, HTX
-
-