Model Number 799-AEB1-00-01 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
Inadequate Osseointegration (2646)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to a faulty capacitor on the smart pneumatic module board.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|