Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: 42532007102, tibia cemented 5 degree stemmed right size e, 64452029.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was discarded by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during an initial knee arthroplasty, the poly would not seat.A different poly was used and successfully implanted.There were no consequences or impact to the patient.
|
|
Manufacturer Narrative
|
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The dhr was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported event is attributed to user error.As stated on the label for part 42522100910, the articular surface is to be used with a size g or h tibia.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|