• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2020
Event Type  malfunction  
Manufacturer Narrative
No remaining sample material was available for further investigation.The investigation could not identify a product problem.The cause of the event could not be determined.This event occurred in (b)(6).
 
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys tsh assay, elecsys ft3 iii, and the elecsys ft4 assay on a cobas 8000 e 801 module and a second e 801 analyzer used for investigation.No alleged erroneous results were reported outside of the laboratory.This medwatch will apply to the tsh assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the ft3 assay and refer to the medwatch with patient identifier (b)(6) for information related to the ft4 assay.The sample was initially tested on the customer's e 801 analyzer on (b)(6) 2020.The sample was repeated on a siemens centaur and an abbott architect analyzer.The sample was also provided for investigation where it was tested on a second e 801 analyzer on (b)(6) 2020.The serial number of the customer's e 801 analyzer is (b)(4).The serial number of the e 801 analyzer used for investigation is (b)(4).Tsh reagent lot number 416233, with an expiration date of october 2020 was used on this analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key9868179
MDR Text Key221477190
Report Number1823260-2020-00813
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTSH
Device Catalogue Number07028091190
Device Lot Number416233
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age56 YR
-
-