Model Number 9735665
Device Problem Output Problem (3005)
Patient Problem No Patient Involvement (2645)
Event Date 03/03/2020
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9735795, serial/lot #: (b)(4). A medtronic representative went to the site to test the equipment. The monitor and cables were replaced. The system then passed a system checkout. The monitor was returned to medtronic for analysis. Analysis revealed that the reported issue was confirmed. When the monitor was powered on with no video animation, the monitor functioned normally, but with video animation the monitor went to a blank/black screen after running for a minute. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database. The cables were also returned to medtronic for analysis. No failures were found. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure. It was reported that the main cart monitor was flickering. There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTH S8 PREMIUM
Type of DeviceINSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9873929
MDR Text Key184730558
Report Number1723170-2020-01047
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/24/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2020
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-