Other relevant device(s) are: product id: 9735795, serial/lot #: (b)(4).
A medtronic representative went to the site to test the equipment.
The monitor and cables were replaced.
The system then passed a system checkout.
The monitor was returned to medtronic for analysis.
Analysis revealed that the reported issue was confirmed.
When the monitor was powered on with no video animation, the monitor functioned normally, but with video animation the monitor went to a blank/black screen after running for a minute.
The hardware investigation found that the reported event was related to a hardware issue.
This issue was documented in a medtronic navigation hardware anomaly tracking database.
The cables were also returned to medtronic for analysis.
No failures were found.
If information is provided in the future, a supplemental report will be issued.
|