Device evaluated by mfr: returned product consisted of an emerge balloon catheter.The balloon was loosely folded with blood in the balloon and the inflation lumen.The device was soaked in a warm water bath for 2 days.Analysis of the tip, balloon, inner/outer shaft, and hypotube included microscopic and visual inspection.Inspection revealed tip damage (flared), and a hole (scrape) in the bottom of the port/exit notch.Inspection of the rest of the device found no other damage or defect.The device was leak tested in the lab, and water was found to be leaking from the hole/scrape in the port weld.The reported rupture was confirmed.
|