Model Number 24970A
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2020
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that when sensing was measured by the programmer at the time of one-week postoperative device check, r sensed waves could not be measured. Rebooting the programmer application did not improve the situation. R sensed wave could be measured over 20mv by using another programmer, therefore the possibility of an anomaly of the application was considered. Because programmer indicated the sensed wave as over 31mv, measurement might have been incapable when the value was excessively high. The programmer remains in use. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC BASE
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9882865
MDR Text Key185252325
Report Number3004593495-2020-00335
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeJA
PMA/PMN NumberP890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 05/06/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received03/26/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/05/2020
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-