• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0302
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 03/06/2020
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report the leak in the steerable guide catheter.It was reported that during preparation of the steerable guide catheter (sgc), it was not able to hold column.The device was not used.There was no clinically significant delay in the procedure and no patient involvement.No additional information was provided.
 
Manufacturer Narrative
The returned device analysis did not confirm the reported leak issue.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.All available information was investigated and a definitive cause for the reported leak could not be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6: device code 1354 removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9884247
MDR Text Key185092341
Report Number2024168-2020-02949
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/19/2020
Device Catalogue NumberSGC0302
Device Lot Number91120U256
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2020
Date Manufacturer Received05/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-