• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN CUP; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN CUP; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Malposition of Device (2616)
Patient Problem No Information (3190)
Event Date 03/10/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Reported event was unable to be confirmed due to limited information received from the customer.X-ray review demonstrated that advanced liner wear of the left hip arthroplasty with eccentric head position as noted and elevated cup abduction angle.There is proximal femoral radiolucency without definite osteolysis although the image is limited.There is subluxation of the femoral head within the cup as noted.Device history record review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2020 - 01107, 0001822565 - 2020 - 01142.
 
Event Description
It was reported the patient underwent an initial total hip arthroplasty on an unknown date.Subsequently the patient was revised due to liner wear.The liner was removed and replaced.Attempts have been made, however no further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN CUP
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key9898320
MDR Text Key189876802
Report Number0001822565-2020-01141
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age69 YR
-
-