• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE DISPOSABLE SETS WT CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE DISPOSABLE SETS WT CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4469C
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/27/2020
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice cassette had fluid leakage due to damage at the tube near the solution bag connector.This was observed during priming for peritoneal dialysis (pd) therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: the device was received for evaluation.A visual inspection with naked eye noted a hole in the heater line tubing at the white spike.Functional testing, including clear passage testing performed and clamp function testing were performed with no issues noted.Leak testing noted leak through a hole in the heater line tubing.The reported condition was verified.The cause of the leak was due to a hole in the heater line tubing.The cause of the hole in the tubing could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE DISPOSABLE SETS WT CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9899235
MDR Text Key185599699
Report Number1416980-2020-01831
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412087375
UDI-Public(01)00085412087375
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K923065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4469C
Device Lot NumberH19G22081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received04/02/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNSPECIFIED PD SOLUTION
-
-