Catalog Number 466P306X
Device Problems Fracture (1260); Failure to Align (2522); Migration (4003)
Patient Problems Embolus (1830); Occlusion (1984); Perforation (2001); Perforation of Vessels (2135)
Event Date 02/10/2019
Event Type  Injury  
Manufacturer Narrative
It was reported that a patient underwent placement of a trapease vena cava filter. The information provided indicated that the filter subsequently malfunctioned and caused perforation, fracture, tilting, migration, occlusion and perforation abutting an adjacent organ. The indication for the filter placement and procedural details have not been provided and there is currently no additional information available for review. The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed. The trapease vena cava filter is indicated for use in the prevention of recurrent pe via percutaneous placement in the inferior vena cava (ivc) for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed or is contraindicated. The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart. Without images or procedural films for review, the reported filter tilt, migration, fracture and perforation could not be confirmed, nor a cause determined. The instructions for use (ifu) states filter fracture is a potential complication of vena cava filters. Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut. Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuosity. A review of the ifu notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters. Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the ifu as such. Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error. Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter. Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters. Occlusion was reported, however with the limited information provided the event could not be further clarified, it is unknown if it was a stenotic or thrombotic event. Blood clots, clotting and/or occlusive thrombosis within the filter and/or the vasculature do not represent a device malfunction. Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics. Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken. Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal brief, a patient underwent placement of a trapease vena cava filter. The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to, perforation, fracture, tilting, migration, occlusion and perforation abutting an adjacent organ.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN TRAPEASE
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60th avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key9900649
MDR Text Key188516842
Report Number1016427-2020-03974
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number466P306X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/30/2020 Patient Sequence Number: 1
-
-