ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem No Code Available (3191)
Event Date 02/27/2020
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
An ophthalmologist reported that during a cataract surgery with an intraocular lens (iol) implant, the trailing haptic was trapped and broken inside the injector during the delivery process.The patient was left aphakic and a secondary iol surgery has been scheduled.Additional information has been requested.
 
Manufacturer Narrative
Additional information has been provided in d.10., h.3., h.6.And h.10.Evaluation summary: the device with the broken haptic was returned in the opened blister tray inside the opened carton.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger has been retracted to mid-nozzle.The reported broken haptic was observed.The haptic was located ahead of the plunger in the nozzle.The distal portion was facing the end of the nozzle.The remainder of the lens was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported event of "haptic trapped and broken inside the injector".A broken haptic was observed in front of the plunger with the distal portion oriented toward the nozzle tip.The plunger was retracted.The plunger position in relation to the broken haptic during advancement cannot be determined.The haptic position may indicate the plunger was not fully advanced.If the plunger is not fully advanced the trailing haptic may not release properly from the device.A qualified viscoelastic was indicated.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9903361
MDR Text Key186607425
Report Number1119421-2020-00560
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberAU00T0
Device Lot Number12746769
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2020
Date Manufacturer Received12/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED DISCOVISC OVD
Patient Outcome(s) Other;
-
-