Manufacturer's investigation conclusion: incidental finding: damaged housing.The reported event of the console not showing flow was not confirmed.The centrimag 1st generation primary console was returned for analysis to the edc and was evaluated and tested.The console booted up as intended and was tested for 24 hours at a speed of 5000 rpm and a flow of ~9 lpm and functioned as intended.The console was tested for another 24 hours at a speed of 1000 rpm with a flow of ~1.5 lpm and functioned as intended.The reported event was not able to be reproduced during testing; however, the console was scrapped due to the damaged housing.The root cause for the reported event was not conclusively determined through this analysis.Centrimag motor instructions for use instructs to always have a back-up centrimag motor available.Centrimag blood pump instructions for use states "always have a spare centrimag blood pump, back-up console and equipment available for change out." centrimag primary console operating manual warns "one back-up console and motor are required in the immediate vicinity of each patient whenever the centrimag blood pump is used.The back-up console must be connected to the back-up motor, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the primary console or primary motor experience a malfunction." no further information was provided.The manufacturer is closing the file on this event.
|