ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0302
Device Problem Positioning Failure (1158)
Patient Problem No Patient Involvement (2645)
Event Date 03/13/2020
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report unable to straighten tip.It was reported during preparation of a steerable guide catheter (sgc), the +/- knob was rotated, but the tip would not curve or straighten.The sgc was not used in the patient and the procedure was successfully completed with a new sgc.There was not patient involvement and no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated, and the reported unable to curve or straighten the guide were not confirmed via return device analysis.A review of the lot history record revealed no manufacturing nonconformities reported to this lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.All the available information was investigated and the cause for the reported unable to curve and straighten steerable guide catheter (sgc) could not be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key9916087
MDR Text Key186362257
Report Number2024168-2020-03260
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/11/2020
Device Catalogue NumberSGC0302
Device Lot Number91010U115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-