The referenced scope was returned to the service center for evaluation.A review of the service history indicates the scope was purchased on november 5, 2018 and has received service once via repairs on august 7, 2019 for an instrument channel leak issue.The evaluation found the bending section cover at the distal end was cut and the internal bending section skeleton was broken.There were excessive broken fibers bundles (black dots) in image.In addition, the distal end cover had burn marks and the bending section cover was cracked.Based on the evaluation results and similar reported events, the cause of the broken bending section issue can potentially be attributed excessive stress/force applied to the device.The operation manual provides warning that states, do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.The ¿instructions for safe use¿ provides several warning statements in an effort to prevent equipment damage and patient injury.¿if any of the following conditions occur during an examination, immediately stop the examination and withdraw the endoscope from the patient.If the up/down angulation control lever does not move.If the angulation control mechanism is not functioning properly.Visually inspect the bending section for no metallic parts protruding from the bending section.Visually inspect the bending section for bends, twist, or other irregularities while the bending section remains straight.Visually inspect the bending section for abnormal bending shape, or other irregularities.Continued use of the endoscope under these conditions could result in patient injury, bleeding, and/or perforation.".
|