MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P6R
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The referenced scope was returned to the service center for evaluation.A review of the service history indicates the scope was purchased on november 5, 2018 and has received service once via repairs on august 7, 2019 for an instrument channel leak issue.The evaluation found the bending section cover at the distal end was cut and the internal bending section skeleton was broken.There were excessive broken fibers bundles (black dots) in image.In addition, the distal end cover had burn marks and the bending section cover was cracked.Based on the evaluation results and similar reported events, the cause of the broken bending section issue can potentially be attributed excessive stress/force applied to the device.The operation manual provides warning that states, do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.The ¿instructions for safe use¿ provides several warning statements in an effort to prevent equipment damage and patient injury.¿if any of the following conditions occur during an examination, immediately stop the examination and withdraw the endoscope from the patient.If the up/down angulation control lever does not move.If the angulation control mechanism is not functioning properly.Visually inspect the bending section for no metallic parts protruding from the bending section.Visually inspect the bending section for bends, twist, or other irregularities while the bending section remains straight.Visually inspect the bending section for abnormal bending shape, or other irregularities.Continued use of the endoscope under these conditions could result in patient injury, bleeding, and/or perforation.".
 
Event Description
The service center was informed that during an unspecified therapeutic procedure, the user facility reported the bending section of the uretero-reno fiberscope was not working properly.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9917841
MDR Text Key208345265
Report Number8010047-2020-01958
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/13/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-